FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 10895541 · Received November 24, 2020

Report

Report Number
9611109-2020-00648
Event Type
Malfunction
Date Received
November 24, 2020
Date of Event
October 26, 2020
Report Date
November 23, 2020
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
PMA / PMN Number
K191402
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT THE CUSTOMER MAINTAINS ITS OWN EQUIPMENT THUS NO LIVANOVA TECHNICIAN WAS DISPATCHED TO THE FACILITY. THE CUSTOMER ORDERED A NEW POWER CABLE TO SOLVE THE PROBLEM AND A LIVANOVA SERVICE TECHNICIAN PROVIDED SUPPORT VIA PHONE. BASED ON THE AVAILABLE INFORMATION, THE REPORTED EVENT WAS MOST LIKELY CAUSED BY A DAMAGED POWER CABLE.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT OF A HEATER-COOLER SYSTEM 3T SHUT-DOWN DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1361682 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-85

Patients

Seq Age Sex Outcome Treatment
1