8015 ALARIS SYSTEM PC UNIT
Report
- Report Number
- 2016493-2020-47517
- Event Type
- Malfunction
- Date Received
- November 24, 2020
- Report Date
- September 12, 2018
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403801518
- PMA / PMN Number
- K091308
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE TSC TROUBLESHOOTING PROCESS. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS NOT RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). PLEASE REFER TO FILE #2018-033277-264732 FROM CAD. (B)(4). NPI SN 13910032 MIKE LENHOF RAN A MANUAL BATTERY CONDITIONING ON PCU8015. THE BATTERY LOGS GAVE LB3 CAPACITY READING OF 5000 WHICH IS OUT SIDE THE SPECIFICATION OF 2700 - 4500 (BASED ON SB 592A). I RECOMMENDED MIKE TO CLEAR THE BATTERY LOGS AND RE-RUN THE MANUAL BATTERY CONDITIONING. OTHERWISE, HE WILL REPLACE A NEW BATTERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1360676 | 8015 ALARIS SYSTEM PC UNIT | PUMP, INFUSION | FRN | CAREFUSION SD | 8015 | 10885403801518 |