NEEDLE NS 18GA 1-1/2IN SB TW
Report
- Report Number
- 3002682307-2020-00356
- Event Type
- Malfunction
- Date Received
- November 24, 2020
- Date of Event
- October 29, 2020
- Report Date
- November 2, 2020
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMI
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. (B)(4). INVESTIGATION SUMMARY: TO AID IN THE INVESTIGATION OF THIS ISSUE, PICTURE SAMPLES WERE PROVIDED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE PICTURE SAMPLES, NO ISSUES WITH THE PRODUCT WERE OBSERVED. THE PRODUCT, MATERIAL REFERENCE NUMBER 304644 AND LOT NUMBER 0251117, IS SHOWN WITHIN THE PICTURES. SINCE NO DEFECT IS PRESENTED, IT HAS BEEN CONSIDERED NOT NECESSARY TO PERFORM THE DHR REVIEW. INVESTIGATION CONCLUSION: AFTER REVISION OF THE PICTURE PROVIDED, WE WOULD LIKE TO INFORM YOU THAT ANY NONCONFORMANCE HAS BEEN OBSERVED. COSTUMER ORDERED THE REFERENCE (B)(4); LOT 0251117, AND THIS IS WHAT WE PROVIDED. THE INFORMATION IN THE BAG, DOES NOT MAKE REFERENCE TO THE FINAL PRODUCT (AS THE BOX DOES), CONSIDERING THE INFORMATION FROM THE BAG BELONGS TO AN INTERMEDIATE PROCESS, (ASSEMBLING PROCESS OF THE NEEDLE). LOT 0245325 IS THE ASSEMBLING LOT OF THE NEEDLES, WHICH IS ALWAYS DIFFERENT TO THE FINAL LOT. ROOT CAUSE DESCRIPTION: THE INFORMATION LISTED ON THE INNER BAGS DOES NOT REFER TO THE FINAL PRODUCT, UNLIKE THE OUTER BOX. THE INFORMATION CAPTURED ON THE BAGS BELONGS TO THE INTERMEDIATE PROCESS, THE ASSEMBLY PROCESS OF THE NEEDLE. LOT NUMBER 0245325 IS THE ASSEMBLING LOT OF THE NEEDLES, WHICH IS A SUB-ASSEMBLY PRODUCT INVOLVED IN THE FINAL PRODUCT. THEREFORE, THESE NUMBERS WOULD DIFFER, WITH THE FINAL PRODUCT LOT NUMBER OF 0251117 AND THE SUB-ASSEMBLY LOT NUMBER OF 0245325. RATIONALE: AS NO DEFECTS WERE IDENTIFIED FOR THIS INCIDENT, FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY.
IT WAS REPORTED THAT NEEDLE NS 18GA 1-1/2IN SB TW HAD INCORRECT LABEL INFORMATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ON THE DELIVERY NOTE AND ON THE OUTSIDE LABEL OF THE CARTON THE SAME BATCH NUMBERS ARE SHOWN. 0251117. HOWEVER, THE INNER BAG SHOWS THE BATCH NUMBER 0245325. THEREFORE, WE HAVE TO BLOCK THE GOODS, BECAUSE A CLEAR TRACEABILITY FOR SFM IS NOT GUARANTEED IN THIS WAY. THE QUICKEST SOLUTION IS THAT WE IMMEDIATELY RECEIVE A CONFIRMATION FROM BD WHICH GOODS ARE IN THE BOX. WHETHER THE NUMBER 0245325 IS CORRECT OR THE NUMBER 0251117.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1354708 | NEEDLE NS 18GA 1-1/2IN SB TW | NEEDLE | FMI | BECTON DICKINSON, S.A. | 0251117 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |