FDA Adverse Event Malfunction Summary report: N

UNI-VENT, MODEL 731

MDR report key: 10883709 · Received November 23, 2020

Report

Report Number
1220908-2020-03776
Event Type
Malfunction
Date Received
November 23, 2020
Report Date
November 6, 2020
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
CBK
UDI-DI
00847946023461
PMA / PMN Number
K111473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION FOR EVALUATION. THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO A FAULTY INTEGRATED CIRCUIT ON THE CENTRAL PROCESSING UNIT (CPU) BOARD. THE CENTRAL PROCESSING UNIT BOARD WAS REPLACED TO REMEDY THE REPORT. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1349427 UNI-VENT, MODEL 731 VENTILATOR CBK ZOLL MEDICAL CORPORATION 799-AEV1-00-01 NA 00847946023461

Patients

Seq Age Sex Outcome Treatment
1