ENDOWRIST;DAVINCI SI
Report
- Report Number
- 2955842-2020-11238
- Event Type
- Malfunction
- Date Received
- November 22, 2020
- Date of Event
- September 25, 2020
- Report Date
- October 28, 2020
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874111475
- PMA / PMN Number
- K050369
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE MARYLAND BIPOLAR FORCEPS INSTRUMENT INVOLVED WITH THIS COMPLAINT AND COMPLETED THE DEVICE EVALUATION. INSPECTION IDENTIFIED INSULATION DAMAGE ON THE CONDUCTOR WIRE. THIS OBSERVATION IS MOST COMMONLY CAUSED BY INSTRUMENT MISHANDLING AND MISUSE, SUCH AS COLLISION OF THE INSTRUMENT WITH A SHARP OBJECT. THE INSTRUMENT WAS TESTED AND PASSED ELECTRICAL CONTINUITY. A REVIEW OF THE SUBMITTED IMAGE WAS PERFORMED BY AN INTUITIVE SURGICAL, INC. (ISI) FAILURE ANALYSIS ENGINEER (FAE). THE FOLLOWING ADDITIONAL INFORMATION WAS PROVIDED: THE IMAGE OF THE MARYLAND BIPOLAR FORCEPS INSTRUMENT IDENTIFIED A LIKELY SMALL AMOUNT OF CONDUCTOR WIRE INSULATION DAMAGE; HOWEVER, THE IMAGE QUALITY WAS POOR AND BLURRY. A CONCLUSIVE DETERMINATION COULD NOT BE CONFIRMED BASED ON THIS ANALYSIS. A REVIEW OF THE SITE'S COMPLAINT HISTORY DOES NOT SHOW ANY ADDITIONAL COMPLAINTS RELATED TO THIS PRODUCT. SYSTEM LOG INVESTIGATION: A REVIEW OF THE INSTRUMENT LOG FOR THE MARYLAND BIPOLAR FORCEPS INSTRUMENT LOT# N10200317 / SEQUENCE (B)(4) ASSOCIATED WITH THIS EVENT HAS BEEN PERFORMED. PER LOGS, THE INSTRUMENT WITH 1 REMAINING USABLE LIFE WAS LAST USED ON (B)(6) 2020 USING SYSTEM SH2006. NO SUBSEQUENT USE WAS RECORDED. THIS COMPLAINT IS BEING REPORTED AS A REPORTABLE MALFUNCTION EVENT DUE TO THE FOLLOWING THE MARYLAND BIPOLAR FORCEPS INSTRUMENTS ARE MULTIPLE-USE ELECTROSURGICAL ENDOSCOPIC INSTRUMENTS WITH A GRASPING TIP TO BE USED IN CONJUNCTION WITH THE DA VINCI SYSTEM AND AN EXTERNAL ELECTROSURGICAL UNIT (ESU). THE INSTRUMENT IS DESIGNED TO PROVIDE ENERGY FROM THE DESIGNATED LOCATION ON THE INSTRUMENT (THE TIP) TO THE PLANNED ANATOMICAL LOCATION WHEN USED AS INTENDED. THE ENERGY IS ACTIVATED BY PRESSING THE DESIGNATED PEDAL ON THE SURGEON SIDE CONSOLE (SSC). IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE MARYLAND BIPOLAR FORCEPS INSTRUMENT WAS FOUND TO HAVE A DAMAGED CABLE. THE USER COMPLETED THE PROCEDURE USING THE BACKUP INSTRUMENT. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE AN MDR REPORTABLE EVENT; HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE FAILURE ANALYSIS IDENTIFIED COMPROMISED CONDUCTOR WIRE INSULATION THAT IS EXPOSED TO PATIENT WITH A PASSED ELECTRICAL CONTINUITY TEST. ALTHOUGH THERE WAS NO ARCING REPORTED OR SEEN THROUGH ANALYSIS, AND THERE WAS NO PATIENT INJURY REPORTED, IF THIS FAILURE WERE TO RECUR, IT COULD RESULT IN AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE MARYLAND BIPOLAR FORCEPS INSTRUMENT WAS FOUND TO HAVE A DAMAGED CABLE. THE USER COMPLETED THE PROCEDURE USING THE BACKUP INSTRUMENT. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1347054 | ENDOWRIST;DAVINCI SI | MARYLAND BIPOLAR FORCEPS | NAY | INTUITIVE SURGICAL, INC | 420172-17 | N10200317 0571 | 00886874111475 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |