FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.5CCM 30GA 8MM BLS 100BX/500 AP

MDR report key: 10881550 · Received November 22, 2020

Report

Report Number
1920898-2020-01606
Event Type
Malfunction
Date Received
November 22, 2020
Date of Event
October 28, 2020
Report Date
March 5, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTOS PROVIDED. (B)(4) PHOTOS OF SHIPPERS AND SHELF CARTONS FOR SYRINGES FROM LOT 0041321 WERE PROVIDED. CUSTOMER REPORTED LOT NUMBER DISAPPEAR. THE PHOTOS WERE REVIEWED, AND IT WAS OBSERVED THAT THE LOT INFORMATION WAS MISSING ON 2 OF THE SHELF CARTONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 0041321. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. BECAUSE THE CARTON WAS CODED BUT IN THE WRONG LOCATION MEANS IT WAS INTRODUCED BACK INTO PRODUCTION BUT NOT PLACED CORRECTLY ON THE TRANSFER BELT BEFORE THE CODING PROCESS.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 0.5CCM 30GA 8MM BLS 100BX/500 AP WAS MISSING LABEL INFORMATION. THIS OCCURRED ON 2 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LOT NUMBER DISAPPEAR.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 0.5CCM 30GA 8MM BLS 100BX/500 AP WAS MISSING LABEL INFORMATION. THIS OCCURRED ON 2 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LOT NUMBER DISAPPEAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1346741 SYRINGE 0.5CCM 30GA 8MM BLS 100BX/500 AP SYRINGE FMF BD MEDICAL - DIABETES CARE 0041321

Patients

Seq Age Sex Outcome Treatment
1