AUTOSTAINER LINK 48
Report
- Report Number
- 2022180-2020-00388
- Event Type
- Malfunction
- Date Received
- November 20, 2020
- Date of Event
- October 23, 2020
- Report Date
- November 20, 2020
- Manufacturer
- DAKO NORTH AMERICA, INC
- Product Code
- KPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ROOT CAUSE: POTENTIAL ALTERATION OF STAINING IN THIS CASE WAS DUE TO IMPROPER OPERATION OF THE BUFFER PUMP. THE PROBLEM WAS SOLVED BY FIELD SERVICE ENGINEER WITH REPLACEMENT OF THE PART. FOLLOWING THE REPLACEMENT, THE INSTRUMENT WAS FULLY OPERATIONAL WITHIN SPECIFICATIONS, WITHOUT ERRORS AND AVAILABLE FOR THE USER. FAILURE MODE DESCRIPTION: FOLLOWING A BUFFER PUMP MALFUNCTION OR IF THE PART STOPS WORKING; THE RESULTING FAILURE MODES COULD OCCUR. BUFFER PUMP IS OUT OF SPECIFICATION DUE TO BROKEN PART, PRIMING FAILURE, DRIPPING, OR TOO MUCH BUFFER BEING APPLIED. THESE FAILURE MODES HAVE THE POTENTIAL TO ALTER STAINING.
CUSTOMER COMPLAINT RECORD REPORTED THE EVENT AS FOLLOWS: BUFFER PUMP PRESSURE TOO WEAK. NO DIRECT OR INDIRECT PATIENT HARM OR USER HARM HAVE BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1343934 | AUTOSTAINER LINK 48 | AUTOSTAINER LINK 48 | KPA | DAKO NORTH AMERICA, INC | AS480 |