FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 10875826 · Received November 20, 2020

Report

Report Number
9611109-2020-00644
Event Type
Malfunction
Date Received
November 20, 2020
Date of Event
October 23, 2020
Report Date
November 16, 2020
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). LIVANOVA FIELD SERVICE TECHNICIAN DISPATCHED TO THE FACILITY CONFIRMED THE ISSUE AND TRACED IT BACK TO A COMPRESSOR DAMAGE. THE UNIT WAS REMOVED FROM SERVICE.

Description of Event or Problem · 1

LIVANOVA RECEIVED A REPORT THAT ON A HEATER COOLER 3T CIRCUIT BREAKER TRIGGERED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1344709 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment