FDA Adverse Event
Malfunction
Summary report: N
HEATER-COOLER SYSTEM 3T
MDR report key: 10875826
·
Received November 20, 2020
Report
- Report Number
- 9611109-2020-00644
- Event Type
- Malfunction
- Date Received
- November 20, 2020
- Date of Event
- October 23, 2020
- Report Date
- November 16, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). LIVANOVA FIELD SERVICE TECHNICIAN DISPATCHED TO THE FACILITY CONFIRMED THE ISSUE AND TRACED IT BACK TO A COMPRESSOR DAMAGE. THE UNIT WAS REMOVED FROM SERVICE.
Description of Event or Problem · 1
LIVANOVA RECEIVED A REPORT THAT ON A HEATER COOLER 3T CIRCUIT BREAKER TRIGGERED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1344709 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |