FDA Adverse Event Injury Summary report: N

MEERA EU WITH AUTO DRIVE

MDR report key: 10874299 · Received November 20, 2020

Report

Report Number
8010652-2020-00025
Event Type
Injury
Date Received
November 20, 2020
Date of Event
November 13, 2020
Report Date
January 22, 2021
Manufacturer
HOLGER ULLRICH
Product Code
FQO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 0

IN THE MEANTIME WE WERE INFORMED BY THE CUSTOMER THAT IN ADDITION TO THE INJURED FOOT OF THE DOCTOR, THE ACTUAL SURGERY OF THE PATIENT WAS DELAYED FOR APPROXIMATELY 1 HOUR DUE TO THE INCIDENT. MANUFACTURER REFERENCE # (B)(4).

Additional Manufacturer Narrative · 1

NO MALFUNCTION OF THE PRODUCT WAS REPORTED BY THE CUSTOMER. THE INCIDENT WAS CAUSED BY A USER ERROR. THE CUSTOMER MADE A FUNCTIONAL CHECK OF THE TABLE AFTER THE PROCEDURE AND NO MALFUNCTION WAS DETECTED. IN THE INSTRUCTIONS FOR USE (IFU) IT IS STATED THE USER HAS TO PAY ATTENTION AND AVOID COLLISIONS WHEN OPERATING THE TABLE. IN CHAPTER 2 OF THE IFU POSSIBLE HAZARDS ARE STATED: "WARNING! RISK OF INJURY! WHEN ADJUSTING AND MOVING THE OR TABLE, THE TABLE TOP OR THE ACCESSORIES, COLLISIONS MAY OCCUR WITH THE PATIENT, BETWEEN THE INDIVIDUAL PRODUCTS OR PARTS POINTING DOWNWARDS. DURING THE ADJUSTMENT PROCEDURES, ALWAYS PAY ATTENTION TO THE OR TABLE AND ACCESSORIES AND AVOID COLLISIONS. ENSURE THAT TUBES, CABLES AND DRAPES ARE NOT TRAPPED." GETINGE-MAQUET GMBH PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.

Description of Event or Problem · 1

THE FOLLOWING APPEARS TO HAVE HAPPENED AND WAS REPORTED TO US. THE TABLE TOP OF THE USED OR TABLE CAN BE LOWERED. IN THIS CASE THE DOWNWARD MOVEMENT OF THE OR TABLE TOP WAS BLOCKED BY AN INSTRUMENT TROLLEY AND THE OR TABLES BASE WAS LIFTED FROM THE GROUND WHICH WAS NOT RECOGNIZED BY THE USER. IT WAS DESCRIBED THAT THE OR TABLE WAS ¿ROCKING¿. THE MEDICAL STAFF STABILIZED THE PATIENTS POSITION ON THE TABLE TOP. AT SOME POINT A SURGEON PUT HER FOOT UNDER THE LIFTED TABLE BASE. WHEN THE TABLE TOP WAS RAISED, THE TABLE BASE WAS LOWERED ON THE DOCTORS FOOT. THE SURGEON INJURED HER FOOD AND ATTENDED THE ACCIDENT AND EMERGENCY DEPARTMENT. SHE HAD TWO BROKEN BONES IN HER FOOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1343368 MEERA EU WITH AUTO DRIVE TABLE, OPERATING-ROOM, AC-POWERED FQO HOLGER ULLRICH 720001B2

Patients

Seq Age Sex Outcome Treatment
1 Other