HF UNIT "ESG-400"
Report
- Report Number
- 9610773-2020-00273
- Event Type
- Malfunction
- Date Received
- November 20, 2020
- Date of Event
- November 12, 2020
- Report Date
- December 21, 2020
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION: D11 - CONCOMITANT MEDICAL PRODUCTS AND THERAPY DATES DEVICE EVALUATION: THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION BUT TO AN OLYMPUS CHINA SALES AND MARKETING (OCSM) REPAIR CENTER. THE EVALUATION/INVESTIGATION AT THE OCSM REPAIR CENTER CONFIRMED THE OCCURRENCE OF ERROR E433. THIS ERROR MESSAGE, WHICH IN THE CASE AT HAND WAS CAUSED BY A DEFECT OF THE GENERATOR BOARD, IS TRIGGERED BY THE GENERATOR¿S SAFETY SYSTEM. IN CASE OF CRITICAL ERRORS, THE SAFETY SYSTEM WILL NOT PERMIT ANY FURTHER USE OF THE GENERATOR UNTIL THE ERROR IS RECTIFIED. THEREFORE, THIS EVENT/INCIDENT WAS ATTRIBUTED TO COMPONENT FAILURE. A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER OF THE HF GENERATOR WITHOUT SHOWING ANY ABNORMALITIES. THE CASE WILL BE CLOSED FROM OLYMPUS SIDE WITH NO FURTHER ACTIONS. THE REPORTED EVENT/INCIDENT WILL BE RECORDED FOR TRENDING AND SURVEILLANCE PURPOSES AND THE USER WILL BE INFORMED ABOUT THE INVESTIGATION RESULTS.
THE SUSPECT MEDICAL DEVICE HAS NOT YET BEEN RETURNED TO OLYMPUS FOR EVALUATION/INVESTIGATION. THEREFORE, THE EXACT CAUSE OF THE USER'S EXPERIENCE AND THE REPORTED PHENOMENON COULD NOT BE DETERMINED AND IS BEING JUDGED AS UNKNOWN. HOWEVER, IF THE SUSPECT MEDICAL DEVICE IS RETURNED FOR EVALUATION/INVESTIGATION OR ADDITIONAL SIGNIFICANT INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED. PLEASE NOTE: THIS REPORT IS BEING SUBMITTED ALTHOUGH THE SUSPECT MEDICAL DEVICE IS NOT MARKETED IN THE USA. HOWEVER, A SIMILAR DEVICE IS MARKETED. MODEL # / CATALOG #: WB91051W; BRAND NAME: HF UNIT "ESG-400"; COMMON DEVICE NAME: HF-GENERATORS; 510(K): K141225; PRODUCT CODE: GEI.
OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE, THE ESG-400 HF-GENERATOR ISSUED ERROR CODE E433. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED USING A SIMILAR DEVICE AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1343510 | HF UNIT "ESG-400" | HF-GENERATORS | GEI | OLYMPUS WINTER & IBE GMBH | WB91051C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | OLYMPUS HF CABLE, BIPOLAR, 4 M (WA00014A) |