FDA Adverse Event Malfunction Summary report: N

HF UNIT "ESG-400"

MDR report key: 10874197 · Received November 20, 2020

Report

Report Number
9610773-2020-00273
Event Type
Malfunction
Date Received
November 20, 2020
Date of Event
November 12, 2020
Report Date
December 21, 2020
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D11 - CONCOMITANT MEDICAL PRODUCTS AND THERAPY DATES DEVICE EVALUATION: THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION BUT TO AN OLYMPUS CHINA SALES AND MARKETING (OCSM) REPAIR CENTER. THE EVALUATION/INVESTIGATION AT THE OCSM REPAIR CENTER CONFIRMED THE OCCURRENCE OF ERROR E433. THIS ERROR MESSAGE, WHICH IN THE CASE AT HAND WAS CAUSED BY A DEFECT OF THE GENERATOR BOARD, IS TRIGGERED BY THE GENERATOR¿S SAFETY SYSTEM. IN CASE OF CRITICAL ERRORS, THE SAFETY SYSTEM WILL NOT PERMIT ANY FURTHER USE OF THE GENERATOR UNTIL THE ERROR IS RECTIFIED. THEREFORE, THIS EVENT/INCIDENT WAS ATTRIBUTED TO COMPONENT FAILURE. A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER OF THE HF GENERATOR WITHOUT SHOWING ANY ABNORMALITIES. THE CASE WILL BE CLOSED FROM OLYMPUS SIDE WITH NO FURTHER ACTIONS. THE REPORTED EVENT/INCIDENT WILL BE RECORDED FOR TRENDING AND SURVEILLANCE PURPOSES AND THE USER WILL BE INFORMED ABOUT THE INVESTIGATION RESULTS.

Additional Manufacturer Narrative · 1

THE SUSPECT MEDICAL DEVICE HAS NOT YET BEEN RETURNED TO OLYMPUS FOR EVALUATION/INVESTIGATION. THEREFORE, THE EXACT CAUSE OF THE USER'S EXPERIENCE AND THE REPORTED PHENOMENON COULD NOT BE DETERMINED AND IS BEING JUDGED AS UNKNOWN. HOWEVER, IF THE SUSPECT MEDICAL DEVICE IS RETURNED FOR EVALUATION/INVESTIGATION OR ADDITIONAL SIGNIFICANT INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED. PLEASE NOTE: THIS REPORT IS BEING SUBMITTED ALTHOUGH THE SUSPECT MEDICAL DEVICE IS NOT MARKETED IN THE USA. HOWEVER, A SIMILAR DEVICE IS MARKETED. MODEL # / CATALOG #: WB91051W; BRAND NAME: HF UNIT "ESG-400"; COMMON DEVICE NAME: HF-GENERATORS; 510(K): K141225; PRODUCT CODE: GEI.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE, THE ESG-400 HF-GENERATOR ISSUED ERROR CODE E433. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED USING A SIMILAR DEVICE AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1343510 HF UNIT "ESG-400" HF-GENERATORS GEI OLYMPUS WINTER & IBE GMBH WB91051C

Patients

Seq Age Sex Outcome Treatment
1 OLYMPUS HF CABLE, BIPOLAR, 4 M (WA00014A)