FDA Adverse Event Malfunction Summary report: N

8015 ALARIS SYSTEM PC UNIT

MDR report key: 10872892 · Received November 19, 2020

Report

Report Number
2016493-2020-43339
Event Type
Malfunction
Date Received
November 19, 2020
Report Date
June 29, 2018
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403801518
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE TSC TROUBLESHOOTING PROCESS. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A TRACKWISE COMPLAINT HISTORY REVIEW WAS COMPLETED, AND IT WAS CONFIRMED THAT THERE WERE ADDITIONAL COMPLAINTS RECEIVED WITH SIMILAR SN (B)(4) FOR THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

(B)(4). CUSTOMER CALLED EXPERIENCING ERROR CODE 110.5010 ON THEIR 8015 AFTER FLASHING SOFTWARE. INFORMED THE CUSTOMER THEY NEED TO FLASH THE PKB SOFTWARE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1337366 8015 ALARIS SYSTEM PC UNIT PUMP, INFUSION FRN CAREFUSION SD 8015 10885403801518

Patients

Seq Age Sex Outcome Treatment
1