HEMOSPRAY ENDOSCOPIC HEMOSTAT
Report
- Report Number
- 1037905-2020-00494
- Event Type
- Malfunction
- Date Received
- November 19, 2020
- Date of Event
- October 26, 2020
- Report Date
- December 21, 2020
- Manufacturer
- COOK ENDOSCOPY
- Product Code
- QAU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
INFORMATION REGARDING SUSPECT MEDICAL DEVICE SECTION COMMON DEVICE NAME: HEMOSTATIC DEVICE FOR INTRALUMINAL GASTROINTESTINAL USE PRODUCT CODE: QAU. INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED COULD NOT CONFIRM THE REPORT. ALL COMPONENTS WERE NOT INCLUDED IN THE RETURN (NO CATHETERS WERE RETURNED), THEREFORE A COMPLETE EVALUATION WAS NOT POSSIBLE. THE DEVICE WAS RETURNED WITH THE ON/OFF SWITCH IN THE "OFF" POSITION. THE RED ACTIVATION KNOB WAS ENGAGED IN THE HANDLE INDICATING ACTIVATION OF THE CARBON DIOXIDE (CO2) CARTRIDGE. THE DEVICE WAS TESTED AS RETURNED AND DID NOT SPRAY AS INTENDED. THE DEVICE DID NOT DISCHARGE WHEN DEACTIVATED AND THE CO2 CARTRIDGE WAS FULLY PUNCTURED. THE LANCE POSITION WAS MEASURED AND FOUND TO BE POSITIONED CORRECTLY. A VISUAL EXAMINATION OF THE O-RING AND LANCE INSIDE THE HANDLE SHOWED BOTH COMPONENTS TO BE POSITIONED CORRECTLY INSIDE THE HANDLE AND THE LANCE TO BE BEVELED. THE INSPECTION OF THE CO2 CARTRIDGE AND REGULATOR (LANCE AND O-RING) CONFIRMS THE DEVICE WAS OF THE CURRENT DESIGN. WHEN TESTED WITH A NEW CO2 CARTRIDGE AND ACTIVATION KNOB, THE DEVICE SPRAYED AS INTENDED. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: THE ROOT CAUSE FOR THE REPORT OF UNABLE TO SPRAY IS UNKNOWN BECAUSE THE DEVICE FUNCTIONED AS INTENDED WHEN USED WITH A NEW CO2 CARTRIDGE. A DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS OF THE RETURNED PRODUCT. HOWEVER, THE REPORT INDICATES THAT THE CATHETERS USED WERE CLOGGED DURING THE PROCEDURE WHICH IS THE MOST LIKELY CAUSE FOR THIS OCCURRENCE. THE CAUSE OF THE CLOGGED CATHETERS IS UNKNOWN. PRIOR TO DISTRIBUTION, ALL HEMOSPRAY ENDOSCOPIC HEMOSTATS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
INFORMATION REGARDING SUSPECT MEDICAL DEVICE: COMMON DEVICE NAME: HEMOSTATIC DEVICE FOR INTRALUMINAL GASTROINTESTINAL USE. PRODUCT CODE: QAU. INITIAL REPORTER OCCUPATION: MATERIAL COORDINATOR. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE REPORT INDICATES THAT THE CATHETERS USED WERE CLOGGED DURING THE PROCEDURE WHICH IS THE MOST LIKELY CAUSE FOR THIS OCCURRENCE. THE CAUSE OF THE CLOGGED CATHETERS IS UNKNOWN. PRIOR TO DISTRIBUTION, ALL HEMOSPRAY ENDOSCOPIC HEMOSTATS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED HEMOSPRAY ENDOSCOPIC HEMOSTAT. THE PHYSICIAN WAS USING HEMOSPRAY DURING THE PROCEDURE AND THE DEVICE STOPPED WORKING. THE HEMOSPRAY WAS ABLE TO SPRAY ENOUGH POWDER TO ACHIEVE HEMOSTASIS. THE DEVICE STOPPED DEPLOYING DURING THE PROCEDURE AFTER THE DEVICE CLOGGED A COUPLE OF TIMES. TROUBLE SHOOTING PROCEDURES WERE APPLIED AND THE CATHETER WAS CLEARED, BUT THE DEVICE WOULD NO LONGER DEPLOY HEMOSPRAY POWDER. OUR ATTEMPTS TO COLLECT ADDITIONAL INFORMATION REGARDING PATIENT OUTCOME WERE UNSUCCESSFUL. WHILE THE COMPLAINANT DID NOT SPECIFY IF THE PATIENT EXPERIENCED ANY ADVERSE EFFECTS OR REQUIRED ADDITIONAL MEDICAL PROCEDURES DUE TO THIS EVENT, THE INFORMATION ABLE TO BE COLLECTED DOES NOT REASONABLY SUGGEST THE PATIENT WAS ADVERSELY IMPACTED. ADDITIONALLY, THE USER STATED THAT HEMOSTASIS WAS ACHIEVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1337867 | HEMOSPRAY ENDOSCOPIC HEMOSTAT | QAU | COOK ENDOSCOPY | W4346143 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ENDOSCOPE, UNKNOWN MAKE OR MODEL |