8300 ALARIS ETCO2 MODULE
Report
- Report Number
- 2016493-2020-41663
- Event Type
- Malfunction
- Date Received
- November 18, 2020
- Report Date
- August 16, 2018
- Manufacturer
- CAREFUSION SD
- Product Code
- CCK
- UDI-DI
- 10885403830013
- PMA / PMN Number
- K031741
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS-USED INFORMATION WERE REVIEWED FOR THE SAP AND TRACK WISE FILES AND FOUND RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE SOURCE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A TRACKWISE COMPLAINT HISTORY REVIEW WAS COMPLETED, AND IT WAS CONFIRMED THAT THERE WERE ADDITIONAL COMPLAINTS RECEIVED WITH SIMILAR SN (B)(4) FOR THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). THE UNIT HAS A BROKEN EXHAUST BUTTON. (B)(4). BROKEN/DAMAGED (BROKEN DOOR)-CUSTOMER STATE BROKEN EXHAUST BUTTON. ERROR 511.2010 (BAD DISPLAY BD). ERROR 511 .2010 CAUSE BY U7 SOCKET. USE CO # 513743 TO FIX PROBLEM (U7 SOCKET). WE DO NOT HAVE TOOL TO REMOVE SOCKET . (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1330126 | 8300 ALARIS ETCO2 MODULE | PUMP, INFUSION | CCK | CAREFUSION SD | 8300 | 10885403830013 |