HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC
Report
- Report Number
- 1037905-2020-00491
- Event Type
- Malfunction
- Date Received
- November 18, 2020
- Date of Event
- March 9, 2020
- Report Date
- November 18, 2020
- Manufacturer
- COOK ENDOSCOPY
- Product Code
- KNQ
- UDI-DI
- 10827002517995
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PK
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION EVALUATION: THE PRODUCT SAID TO BE INVOLVED WAS RETURNED IN AN OPEN POUCH FROM THE LOT NUMBER PROVIDED IN THE REPORT. THE LABEL MATCHES THE PRODUCT RETURNED. OUR EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THE REPORT. A VISUAL EVALUATION OF THE PRODUCT RECEIVED CONFIRMED THAT THE BLUE CATHETER HAD BROKEN AT THE 130 CM LOCATION AS MEASURED FROM THE DISTAL TIP OF THE BALLOON. NO OTHER KINKS, CRACKS OR ANY OTHER DAMAGE WAS IDENTIFIED ON THE DEVICE. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A CORRECTIVE ACTION HAS BEEN INITIATED TO REDUCE OCCURRENCES OF CATHETER CRACKING, SPLITTING OR BREAKING FOR HBD-W DEVICES. REFERENCE CAPA 261610. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTIONS. PRIOR TO DISTRIBUTION, ALL HERCULES 3 STAGE WIRE GUIDED BALLOONS ESOPHAGEAL-PYLORIC-COLONIC ARE SUBJECTED TO A LEAK TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING AN ESOPHAGEAL STRICTURE DILATION (OGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC AND FOUND THAT THE BALLOON WOULD NOT INFLATE ON DILATION. THE DEVICE WAS RETURNED FOR INVESTIGATION ON 11/02/2020 AND THE BLUE CATHETER WAS BROKEN AT 116.9 CM DISTAL FROM THE HANDLE [SUBJECT OF REPORT]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENTS BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1324751 | HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC | KNQ, DILATOR, ESOPHAGEAL | KNQ | COOK ENDOSCOPY | W4063374 | 10827002517995 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |