VACUETTE 4ML K2E K2EDTA NR LIL/S 13X75
Report
- Report Number
- 1125230-2017-00016
- Event Type
- Malfunction
- Date Received
- November 16, 2020
- Report Date
- November 16, 2020
- Manufacturer
- GREINER BIO-ONE NA INC.
- Product Code
- JKA
- PMA / PMN Number
- K971236
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
COMPLAINT (B)(4). RECEIVED 50PC 454209/B17063C8 FOR EVALUATION. WE HAVE NO FURTHER COMPLAINTS ON THE MATERIAL/BATCH. WE HAVE NO FURTHER INVENTORY OF THE MATERIAL/BATCH. SAMPLES WERE TESTED, ACCORDING TO GBO STANDARD TESTING PROCEDURES, WITH REGARDS TO CORRECT ASSEMBLY, LEVEL MARK, FILLING LEVEL, AND DRAW VOLUME ACCORDING TO ISO 6710 'SINGLE USE CONTAINERS FOR VENOUS BLOOD SPECIMEN COLLECTION' AND CLSI GP39-A6 REGULATIONS FOR 'EVACUATED TUBES AND ADDITIVES FOR BLOOD SPECIMEN COLLECTION'. TUBES WERE VERIFIED TO BE CORRECTLY ASSEMBLED AND HAD THE CORRECT FILL GUIDE LINE POSITION. GREINER FILL MARK PROVIDES A VISUAL CONTROL OPPORTUNITY FOR THE PHLEBOTOMIST AND FOR THE LAB PERSONNEL TO CHECK FOR PROPER VOLUME COLLECTION OF SPECIMEN. BOTH STANDARDS SPECIFY THE DRAW VOLUME TO BE WITHIN +/- 10% RANGE OF THE NOMINAL FILL VOLUME. NO VISUAL DEVIATION IN FILL VOLUME, SPORADIC LOW OR HIGH FILL, COULD BE OBSERVED IN THE TESTED SAMPLES. ALL TUBES TESTED FILLED WITHIN THE +/-10% TOLERANCE RANGE. NO SHORT FILLS COULD BE OBSERVED ON THE SAMPLES. THE COMPLAINT COULD NOT BE CONFIRMED.
CUSTOMER STATES TUBES ARE UNDERFILLING. THIS HAS OCCURRED ONCE WITH TWO TUBES FROM THIS LOT; 1 PHLEBOTOMIST, USING A STRAIGHT NEEDLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1314540 | VACUETTE 4ML K2E K2EDTA NR LIL/S 13X75 | EVACUATED BLOOD COLLECTION TUBE | JKA | GREINER BIO-ONE NA INC. | 454209 | B17063C8 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |