FDA Adverse Event Malfunction Summary report: N

BC UNIVERSAL-QLTD-M

MDR report key: 10843653 · Received November 16, 2020

Report

Report Number
3007802293-2020-00025
Event Type
Malfunction
Date Received
November 16, 2020
Date of Event
July 18, 2020
Report Date
November 16, 2020
Manufacturer
HANDICARE USA INC.
Product Code
FSA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 1

THE SLING INVOLVED IN THE INCIDENT WAS DISCARDED AND NOT RETURNED TO THE MANUFACTURER. PHOTOS OF THE SLING AND DISCUSSION WITH THE USER FACILITY WAS USED TO INVESTIGATE THE ISSUE. THE USER MANUAL DESCRIBES THE LIFE CYCLE OF A HANDICARE SLING TO BE 1 WASH PER WEEK FOR UP TO 4 YEARS WHICH IS EQUIVALENT TO ~208 WASH/DRY CYCLES. THE SLING INVOLVED IN THE INCIDENT IS NEARING THE END OF ITS LIFESPAN AS IT HAS BEEN USED AND WASHED MUCH MORE FREQUENTLY. THE STRAP FAILURE IS DUE TO LONG TERM USE FATIGUE. THE INSTRUCTIONS PROVIDED WITH EACH SLING DIRECT THE USER TO VISUALLY INSPECT THE SLING AND STRAPS PRIOR TO AND DURING A PATIENT TRANSFER. THESE INSTRUCTIONS ARE DEEMED SUFFICIENT AND NO CORRECTIVE ACTION IS NECESSARY. ALL SLINGS AT THE FACILITY WERE INSPECTED AND REPLACED AS NEEDED.

Description of Event or Problem · 1

AT THE BEGINNING OF A BED-WHEELCHAIR TRANSFER WITH A CEILING LIFT, THE LEG STRAP OF THE SLING BROKE. THE RESIDENT WAS NOT INJURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1312849 BC UNIVERSAL-QLTD-M PATIENT SLING FSA HANDICARE USA INC. 537320

Patients

Seq Age Sex Outcome Treatment
1