FDA Adverse Event
Malfunction
Summary report: N
FREECURVE
MDR report key: 10843045
·
Received November 16, 2020
Report
- Report Number
- 3013423626-2020-00023
- Event Type
- Malfunction
- Date Received
- November 16, 2020
- Date of Event
- October 29, 2020
- Report Date
- November 16, 2020
- Manufacturer
- HANDICARE STAIRLIFTS B.V
- Product Code
- PCD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 1
HANDICARE, THE IMPORTER, NOTIFIED THE MANUFACTURER, HANDICARE STAIRLIFTS BV, OF THE INCIDENT ON (B)(6) 2020. THIS UNIT HAS NOT YET BEEN RETURNED. WHEN RECEIVED, IT WILL BE SENT TO HANDICARE STAIRLIFTS BV FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH THE FINDINGS OF THE INVESTIGATION.
Description of Event or Problem · 1
DURING THE INITIAL TEST RIDE BY THE DEALER PRIOR TO SIGNING OFF THE LIFT TO THE CUSTOMER, THE CHAIR TILTED MORE THAN 90 DEGREES. THE DEALER WAS RIDING IN THE UP DIRECTION AT THE TIME, AND WAS ABLE TO CATCH HIMSELF TO PREVENT A FALL. THE DEALER REPLACED THE POWERPACK WITH A NEW UNIT WHICH SUCCESSFULLY PASSED ALL TESTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1311558 | FREECURVE | CHAIR LIFT | PCD | HANDICARE STAIRLIFTS B.V | C0020091 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|