FDA Adverse Event Malfunction Summary report: N

MASTERSCREEN

MDR report key: 10841278 · Received November 14, 2020

Report

Report Number
9615102-2020-00116
Event Type
Malfunction
Date Received
November 14, 2020
Date of Event
May 15, 2020
Report Date
May 15, 2020
Manufacturer
VYAIRE MEDICAL
Product Code
BZC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). VYAIRE MEDICAL DETERMINED THAT THE ROOT CAUSE OF THE BURNING SMELL IS A DEFECTIVE ELECTRICAL OF MS SYSTEMBOX 115V (110V/125V), 430MM. TO ADDRESS THE REPORTED ISSUE, THE COMPONENT HAS BEEN REPLACED. THE UNIT WAS TESTED AND PASSED ALL TESTS.

Description of Event or Problem · 1

IT WAS REPORTED TO VYAIRE MEDICAL THAT WHEN UNIT WAS POWERED ON THERE WAS A DISTINCTIVE SMELL OF BURNING COMING FROM MS PFT RT SES WITH PC. THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1306365 MASTERSCREEN CALCULATOR, PULMONARY FUNCTION DATA BZC VYAIRE MEDICAL MS PFT RT SES WITH PC