FDA Adverse Event
Malfunction
Summary report: N
MASTERSCREEN
MDR report key: 10841278
·
Received November 14, 2020
Report
- Report Number
- 9615102-2020-00116
- Event Type
- Malfunction
- Date Received
- November 14, 2020
- Date of Event
- May 15, 2020
- Report Date
- May 15, 2020
- Manufacturer
- VYAIRE MEDICAL
- Product Code
- BZC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). VYAIRE MEDICAL DETERMINED THAT THE ROOT CAUSE OF THE BURNING SMELL IS A DEFECTIVE ELECTRICAL OF MS SYSTEMBOX 115V (110V/125V), 430MM. TO ADDRESS THE REPORTED ISSUE, THE COMPONENT HAS BEEN REPLACED. THE UNIT WAS TESTED AND PASSED ALL TESTS.
Description of Event or Problem · 1
IT WAS REPORTED TO VYAIRE MEDICAL THAT WHEN UNIT WAS POWERED ON THERE WAS A DISTINCTIVE SMELL OF BURNING COMING FROM MS PFT RT SES WITH PC. THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1306365 | MASTERSCREEN | CALCULATOR, PULMONARY FUNCTION DATA | BZC | VYAIRE MEDICAL | MS PFT RT SES WITH PC |