FDA Adverse Event Malfunction Summary report: N

PEN NDL 32G 6MM 3B TW 70CT JPN

MDR report key: 10840854 · Received November 13, 2020

Report

Report Number
9616656-2020-01138
Event Type
Malfunction
Date Received
November 13, 2020
Date of Event
October 21, 2020
Report Date
February 26, 2021
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 11/4/2020. H.6. INVESTIGATION: TWO OPEN 32G X 6MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 9121516, CAT. NO. 320738. A CLOG TESTING AS PER TP700483 WAS CARRIED OUT ON THE RETURNED SAMPLES AND NO ISSUE WAS OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. NO ISSUES WERE OBSERVED WITH THE RETURNED SAMPLES THEREFORE NO ROOT CAUSE CAN BE IDENTIFIED H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PEN NDL 32G 6MM 3B TW 70CT JPN WAS UNABLE TO DELIVER MEDICATION DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE CUSTOMER REPORTED ABOUT NEEDLE CLOG."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A PEN NDL 32G 6MM 3B TW 70CT JPN WAS UNABLE TO DELIVER MEDICATION DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE CUSTOMER REPORTED ABOUT NEEDLE CLOG."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1305374 PEN NDL 32G 6MM 3B TW 70CT JPN HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 9121516

Patients

Seq Age Sex Outcome Treatment
1