PEN NDL 32G 6MM 3B TW 70CT JPN
Report
- Report Number
- 9616656-2020-01138
- Event Type
- Malfunction
- Date Received
- November 13, 2020
- Date of Event
- October 21, 2020
- Report Date
- February 26, 2021
- Manufacturer
- BECTON DICKINSON AND CO.
- Product Code
- FMI
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 11/4/2020. H.6. INVESTIGATION: TWO OPEN 32G X 6MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 9121516, CAT. NO. 320738. A CLOG TESTING AS PER TP700483 WAS CARRIED OUT ON THE RETURNED SAMPLES AND NO ISSUE WAS OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. NO ISSUES WERE OBSERVED WITH THE RETURNED SAMPLES THEREFORE NO ROOT CAUSE CAN BE IDENTIFIED H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT A PEN NDL 32G 6MM 3B TW 70CT JPN WAS UNABLE TO DELIVER MEDICATION DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE CUSTOMER REPORTED ABOUT NEEDLE CLOG."
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT A PEN NDL 32G 6MM 3B TW 70CT JPN WAS UNABLE TO DELIVER MEDICATION DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE CUSTOMER REPORTED ABOUT NEEDLE CLOG."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1305374 | PEN NDL 32G 6MM 3B TW 70CT JPN | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON DICKINSON AND CO. | 9121516 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |