FDA Adverse Event
Malfunction
Summary report: N
PRECISION XTRA/OPTIUM
MDR report key: 1083750
·
Received July 24, 2008
Report
- Report Number
- 2954323-2008-02327
- Event Type
- Malfunction
- Date Received
- July 24, 2008
- Date of Event
- July 3, 2008
- Report Date
- July 24, 2008
- Manufacturer
- ABBOTT DIABETES CARE LIMITED UK
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER'S PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE FILED ONCE INVESTIGATION RESULTS ARE AVAILABLE.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR PRECISION XTRA BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 340 MG/DL AND 90 MG/DL WITHIN 10 MINUTES. ALL TESTS WERE PERFORMED ON THE FINGER. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION XTRA/OPTIUM | BLOOD GLUCOSE METER | NBW | ABBOTT DIABETES CARE LIMITED UK | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |