FDA Adverse Event
Malfunction
Summary report: N
GE OEC STENOSCOPE
MDR report key: 1083743
·
Received July 29, 2008
Report
- Report Number
- 9617766-2008-01229
- Event Type
- Malfunction
- Date Received
- July 29, 2008
- Date of Event
- July 7, 2008
- Report Date
- July 28, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HIGH VOLTAGE CONNECTOR WAS TIGHTENED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE STENOSCOPE SYSTEM HAD LINES IN THE IMAGES. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC STENOSCOPE | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | NA | NA |