ABACUS V2.0 TPN CALCULATING SOFTWARE
Report
- Report Number
- 1419106-2008-00013
- Event Type
- Other
- Date Received
- July 24, 2008
- Date of Event
- June 25, 2008
- Report Date
- June 25, 2008
- Manufacturer
- BAXA CORPORATION
- Product Code
- LNX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHARMACIST
Narratives
THE CUSTOMER'S SUBSEQUENT INVESTIGATION REVEALED THAT THE PHARMACIST, WHO ENTERED THE SUBJECT ORDER WITHIN THE ABACUS CALCULATING SOFTWARE, MISTAKENLY ENTERED A VALUE OF 3ML/KG OF PEDIATRIC TRACE ELEMENTS IN ADDITION TO 3ML OF ADULT TRACE ELEMENTS. THE FACILITIES PROCEDURE STATES THAT PATIENTS SHOULD RECEIVE ADULT TRACE ELEMENTS AND NO PEDIATRIC TRACE ELEMENTS. PHARMACIST VERIFICATION OF THE ORDERED INGREDIENTS FAILED TO IDENTIFY THE ADDITIONAL 84ML OF PEDIATRIC TRACE ELEMENTS. WITH THE ASSISTANCE OF BAXA TECHNICAL SUPPORT, THE IMPORTANCE OF ABACUS SOFTWARE WARNING LIMITS AND INGREDIENTS DOSING RULES WERE ADDRESSED WITH THE CUSTOMER AND A NEW CATASTROPHIC WARNING LIMIT WAS CREATED FOR PEDIATRIC TRACE ELEMENTS FOR ALL PEDIATRIC AND NEONATAL PATIENTS. THIS WARNING LIMIT WILL NOTIFY THE USER WHEN THE VALUE FOR PEDIATRIC TRACE ELEMENTS HAS EXCEEDED THE MAXIMUM VALUE OF 5ML. INVESTIGATION HAS CONFIRMED THAT THE ABACUS CALCULATING SOFTWARE PERFORMED AS DESIGNED AND DELIVERED THE VOLUME OF PEDIATRIC TRACE ELEMENTS AS ORDERED.
BAXA WAS NOTIFIED ON 06/25/2008 THAT A PATIENT HAD RECEIVED A TPN BAG CONTAINING A HIGHER-THAN-DESIRED QUANTITY OF TRACE ELEMENTS. THE BAG WAS INFUSED FOR 18 HOURS AND THEN REMOVED ONCE THE ISSUE WAS DISCOVERED. THE PATIENT'S BLOOD WORK INDICATED HIGHER-THAN-EXPECTED LEVELS OF ZINC AND CHROMIUM. HOWEVER, NO ADDITIONAL MEDICAL INTERVENTION WAS NEEDED. THE PATIENT IS IN STABLE CONDITION AND DID NOT EXPERIENCE ANY ADVERSE SYMPTOMS AS A RESULT OF THE INFUSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ABACUS V2.0 TPN CALCULATING SOFTWARE | ABACUS MULTI WORKSTATION | LNX | BAXA CORPORATION | 8300-0047 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 6 YR | Required Intervention |