FDA Adverse Event Injury Summary report: N

CONTOUR CURVED CUTTER STAPLER

MDR report key: 1083272 · Received July 29, 2008

Report

Report Number
3005075853-2008-00753
Event Type
Injury
Date Received
July 29, 2008
Date of Event
June 6, 2008
Report Date
July 7, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
GDW
PMA / PMN Number
K040038
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH THE ORIGINAL RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE RELOAD DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN LAR PROCEDURE, AFTER FIRING THE INSTRUMENT IT WAS OBSERVED THAT ONE STAPLELINE WAS MISSING. TO FINISH THE PROCEDURE, THE SURGEON HAD TO HAND SUTURE AND CREATE A TEMPORARY STOMA. NO ADVERSE PT CONSEQUENCES WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR CURVED CUTTER STAPLER GDW ETHICON ENDO-SURGERY, LLC NA D4JW0J

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention