FDA Adverse Event Malfunction Summary report: N

SYRINGE S2 5ML 23GA 1-1/4IN BD CHINA

MDR report key: 10832535 · Received November 12, 2020

Report

Report Number
3002682307-2020-00340
Event Type
Malfunction
Date Received
November 12, 2020
Date of Event
September 30, 2020
Report Date
January 11, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR PROVIDED LOT NUMBER 1903122 AND THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED INCIDENT. AS SAMPLES WERE UNAVAILABLE FOR RETURN, OUR QUALITY TEAM OBTAINED TWENTY RETAINED SAMPLES OF THE SAME LOT NUMBER FROM THE MANUFACTURING FACILITY FOR FURTHER EVALUATION. THE RETAINED SAMPLES WERE EXAMINED AND NO INK PARTICLES OR OTHER DEFECTS WERE OBSERVED WITH THE SYRINGES. BASED ON THE INVESTIGATION FINDINGS, THIS EXACT CAUSE COULD NOT BE CONFIRMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE S2 5ML 23GA 1-1/4IN BD CHINA HAD FOREIGN MATTER INSIDE THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 2020-09-30, IT IS FOUND THAT THERE IS BLACK INK INSIDE THE UNOPENED SYRINGE BEFORE USE, SO IT SHOULD BE STOPPED AND REPLACED WITH A NEW SYRINGE.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE S2 5ML 23GA 1-1/4IN BD (B)(4) HAD FOREIGN MATTER INSIDE THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (B)(6) 2020, IT IS FOUND THAT THERE IS BLACK INK INSIDE THE UNOPENED SYRINGE BEFORE USE, SO IT SHOULD BE STOPPED AND REPLACED WITH A NEW SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1298952 SYRINGE S2 5ML 23GA 1-1/4IN BD CHINA SYRINGE FMF BECTON DICKINSON, S.A. 1903122

Patients

Seq Age Sex Outcome Treatment
1