PLUM XLMD PUMP

MDR report key: 1083240 · Received July 24, 2008

Report

Report Number
2921482-2008-00228
Event Type
Malfunction
Date Received
July 24, 2008
Date of Event
May 3, 2008
Report Date
June 26, 2008
Manufacturer
HOSPIRA, INC.
Product Code
FRN
PMA / PMN Number
K010924
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE DID NOT SOUND AN AUDIBLE ALARM DURING AN ALARM CONDITION WHILE ON THE LOW VOLUME SETTING. THE PUMP WAS RETURNED TO THE BIOMEDICAL ENGINEERING DEPARTMENT WITH A NOTE THAT STATED, "SOUND DOES NOT WORK ON LOW." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION WHILE ON THE LOW VOLUME SETTING. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.