RPM DIAGNOSTIC REFERENCE CATHETER
Report
- Report Number
- 2953184-2008-00029
- Event Type
- Malfunction
- Date Received
- July 24, 2008
- Date of Event
- May 28, 2008
- Report Date
- June 24, 2008
- Manufacturer
- BOSTON SCIENTIFIC, EP TECHNOLOGIES, INC.
- Product Code
- DRF
- PMA / PMN Number
- K992912
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT OF THE DEVICE ON 06/24/2008, THE DEVICE WAS VISUALLY INSPECTED. THE PRIMACORE ON THE 4TH TRANSDUCER WAS FOUND TORN ON THE DISTAL ASSEMBLY, EXPOSING THE POLYIMIDE, TORN PEBAX TUBING AND ELECTRICAL WIRES. THE ELECTRICAL WIRES APPEARED TO BE INTACT (NO WIRES PROTRUDING OUT). AN ELECTRICAL VERIFICATION TEST WAS PERFORMED PRIOR TO DECONTAMINATION AND ALL TRANSDUCERS CAPACITANCE AND RESISTANCE VALUES VERIFIED NORMAL. THE CATHETER WAS DECONTAMINATED AND THE DISTAL TUBING WAS DISSECTED. DURING VISUAL INSPECTION WITH A 10X MICROSCOPE, THE WASHER NEXT TO THE 4TH TRANSDUCER WAS FOUND PUSHED INSIDE THE DISTAL TUBING. A SMALL AMOUNT OF GLUE WAS PRESENT INSIDE THE DISTAL TUBING AND ON THE POLYIMIDE TUBING NEXT TO THE 4TH TRANSDUCER. DURING DHR REVIEW, THERE WERE NO MANUFACTURING ISSUES IDENTIFIED. THE MOST PROBABLE ROOT CAUSE MAY BE DUE TO POOR ADHESIVE BONDING TECHNIQUE DURING MANUFACTURING. EITHER THE ADHESIVE APPLIED TO HOLD THE ASSEMBLY TOGETHER WAS NOT SUFFICIENT OR NOT ENOUGH TIME WAS TAKEN FOR ADHESIVE SETTING. COMPLAINT HISTORY REVIEW FOUND NO SIMILAR COMPLAINTS FOR THIS PRODUCT FAMILY IN THE LAST 24 MONTHS. NO CORRECTIVE ACTION WILL BE TAKEN AS THE PRODUCTION OF THIS PRODUCT WAS DISCONTINUED IN DECEMBER 2007.
THE FOLLOWING INFORMATION WAS REPORTED TO BSC IN 2008. DURING THE PROCEDURE, THE CATHETER WAS INTRODUCED INTO A 7F SHEATH AND NO SIGNAL APPEARED. WHEN THE CATHETER WAS WITHDRAWN, IT WAS DAMAGED. ANOTHER DEVICE WAS USED AND THE PROCEDURE COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RPM DIAGNOSTIC REFERENCE CATHETER | DRF | BOSTON SCIENTIFIC, EP TECHNOLOGIES, INC. | 1500 | 9525523 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | CORDIS AVANT 7 FRENCH SHEATH |