FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL

MDR report key: 1083156 · Received July 24, 2008

Report

Report Number
2023826-2008-00991
Event Type
Malfunction
Date Received
July 24, 2008
Date of Event
July 1, 2008
Report Date
July 2, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

RESULTS: VISUAL INSPECTION OF THE RETURNED PROD FOUND A PIECE OF THE LENS OPTIC AND ONE HAPTIC TORN OFF. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. IT SHOULD BE NOTED THAT THE INJECTOR, CARTRIDGE, AND FOAM TIP PLUNGER WERE NOT RETURNED FOR EVAL. CONCLUSIONS: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CANNOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYS ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYS ISSUES, ALL STAGES IN THE MFG OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

THE RPTR STATED THE SURGEON INSERTED A CC4204BF COLLAMER SINGLE PIECE LENS AND THE LENS TORE. THERE WAS NO PT INJURY, NO ENLARGED INCISION OR SUTURES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL CO. CC4204BF NA

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE MODEL SFC-25 FP - LOT# UNK| INJECTOR MODEL INDIGO-P -LOT# UNK| FOAM TIP PLUNGER MODEL FTP INDIGO - LOT# UNK| VISCOELASTIC MODEL STAARVISC II - LOT # UNK