FDA Adverse Event Malfunction Summary report: N

MOBILE INFANT WARMER

MDR report key: 1083137 · Received July 24, 2008

Report

Report Number
9611451-2008-00477
Event Type
Malfunction
Date Received
July 24, 2008
Report Date
July 8, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
FMT
PMA / PMN Number
K970432
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: AN INITIAL INVESTIGATION WAS CARRIED OUT BASED ON THE EVENT DESCRIPTION AND RESULTS OF PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS, AS THE COMPLAINT DEVICE HAS NOT BEEN RETURNED FOR INVESTIGATION. RESULTS: CRACKING OF THE INFANT WARMER HEAD IS USUALLY DUE TO THE USE OF INCOMPATIBLE CLEANING SOLUTIONS AND DOING THE CLEANING BEFORE THE WARMER HEAD HAS BEEN COOLED. IT WAS REPORTED THAT THE HOSPITAL WAS USING A MIXTURE OF PRECISE QTB, ALKYL DIMETHYL BENZYL, AMMONIUM CHLORIDE AND ETHYL BENZYL BENZOYL IN CLEANING THE INFANT WARMER HEAD. THE OPERATING MANUAL STATES THAT CLEANING OF ALL PARTS OF THE INFANT WARMER AND ACCESSORIES SHOULD BE DONE AT NORMAL ROOM TEMPERATURE, AROUND 23 DEGREES C. THE DEVICE IS MADE UP OF POLYCARBONATE PLASTICS AND CLEANING SOLUTIONS CONTAINING AROMATIC HYDROCARBONS, AMMONIA, AMINES OR AQUEOUS SOLUTIONS ARE NOT RECOMMENDED AS THESE MAY CAUSE CHEMICAL REACTIONS TO THE PLASTIC MATERIAL. ALL SURFACES OF THE WARMER HEAD MAY BE CLEANED WITH DETERGENT BASED SOLUTION WITH MAXIMUM CONCENTRATION OF 2% IN WATER. CONCLUSION: CRACKING OF THE INFANT WARMER HEAD, DUE TO THE USE OF NON-COMPATIBLE CLEANING MATERIALS, IS A KNOWN PROBLEM. WE HAVE AN OPEN CAPA TO EXAMINE FURTHER COMPATIBLE CLEANING SOLUTIONS.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THEIR IW910JEU MOBILE INFANT WARMER HEAD HAD CRACKED AS THEY WERE USING THE WRONG CLEANING PRODUCT. THEY WERE USING A MIXTURE OF PRECISE QTB, ALKYL DIMETHYL BENZYL, AMMONIUM CHLORIDE AND ETHYL BENZYL BENZOYL IN CLEANING THE INFANT WARMER HEAD. NO PT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MOBILE INFANT WARMER FMT FISHER & PAYKEL HEALTHCARE, LTD. IW910JEU

Patients

Seq Age Sex Outcome Treatment
1