FDA Adverse Event Malfunction Summary report: N

ARCHITECT I2000SR ANALYZER

MDR report key: 1083126 · Received July 24, 2008

Report

Report Number
1628664-2008-00179
Event Type
Malfunction
Date Received
July 24, 2008
Report Date
July 2, 2008
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
PMA / PMN Number
K983212
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATES THAT ONE PATIENT WITH CLINICAL FINDINGS SUGGESTING A HEPATITIS A INFECTION GENERATED A NON-REACTIVE ARCHITECT HAVAB-IGM INDEX RESULT OF 0.39. THE PATIENT IS JAUNDICED WITH VERY ELEVATED AST AND ALT RESULTS. THE PATIENT WAS HOSPITALIZED, WHERE TESTING WAS REACTIVE FOR HABAB-IGM. A SECOND SAMPLE WAS DRAWN AND GENERATED AN ARCHITECT HAVAB-IGM INDEX RESULT OF 0.66 (NON-REACTIVE). THIS SECOND SAMPLE GENERATED A REACTIVE RESULT ON THE AXSYM AND DIASORIN PLATFORMS. THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT I2000SR ANALYZER CHEMILUMINESCENT MICROPARTICLE IMMUNOASSAY ANALYZER JJE ABBOTT MANUFACTURING, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR ARCHITECT HAVAB IGM ASSAY| AND 60318HN00| LOT#: 59153HN00