ENDOPATH XCEL BLADELESS TROCAR WITH STABILITY SLEEVE - 12MM DIAMETER, 100MM LENG
Report
- Report Number
- 1527736-2008-03883
- Event Type
- Malfunction
- Date Received
- July 24, 2008
- Date of Event
- May 14, 2008
- Report Date
- June 3, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, INC. (CINCINNATI)
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
INSTRUMENT B: EXP DATE: 02/05/2012; MFR DATE: 03/05/2007. INSTRUMENT C AND D: EXP DATE: 03/01/2013; MFR DATE: 04/01/2008. LEAKED, FRACTURED CLEAR LENS. EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT A WAS LEAK TESTED AND FAILED. ADDITIONALLY THE LENS OF THE OBTURATOR WAS NOTED TO BE CRACKED. NO ANOMALIES WERE NOTED WITH THE UNIVERSAL SEAL ASSEMBLY, IT WAS NOTED TO BE PROPERLY ASSEMBLED. THE ANALYSIS RES ULTS FOUND THAT THE INSTRUMENT B WAS FUNCTIONALLY LEAK TESTED AND FAILED. ADDITIONALLY THE LENS OF THE OBTURATOR WAS NOTED TO BE CRACKED. NO ANOMALIES WERE NOTED WITH THE UNIVERSAL SEAL ASSEMBLY, IT WAS NOTED TO BE PROPERLY ASSEMBLED. THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT C WAS FUNCTIONALLY LEAK TESTED AND FAILED. ADDITIONALLY THE LENS OF THE OBTURATOR WAS NOTED TO BE CRACKED. NO ANOMALIES WERE NOTED WITH THE UNIVERSAL SEAL ASSEMBLY, IT WAS NOTED TO BE PROPERLY ASSEMBLED. THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT D WAS FUNCTIONALLY LEAK TESTED AND FAILED. ADDITIONALLY THE LENS OF THE OBTURATOR WAS NOTED TO BE CRACKED. NO ANOMALIES WERE NOTED WITH THE UNIVERSAL SEAL ASSEMBLY, IT WAS NOTED TO BE PROPERLY ASSEMBLED. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING LAP GASTRIC BYPASS SURGERY THERE WERE 5 TROCARS USED THAT LEAKED DURING THE CASE. THEY WERE SEPARATED FROM WHERE THE HOUSING IS ATTACHED TO THE CANNULA. THEY WERE ABLE TO COMPLETE THE CASE WITH THE 5TH TROCAR WITH NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH XCEL BLADELESS TROCAR WITH STABILITY SLEEVE - 12MM DIAMETER, 100MM LENG | GCJ | ETHICON ENDO-SURGERY, INC. (CINCINNATI) | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |