FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1083043 · Received July 23, 2008

Report

Report Number
1527736-2008-03876
Event Type
Malfunction
Date Received
July 23, 2008
Report Date
June 9, 2008
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE HAND PIECE WAS TESTED ON A GENERATOR AND GAVE ERROR CODE 4. THE HAND PIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE DEVICE BECAME HOT. UNK HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE LFL ETHICON ENDO-SURGERY, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR| HANDPIECE