C4120, 38CM LATIS GRASPER CART, 10/BX
Report
- Report Number
- 2027111-2020-00616
- Event Type
- Malfunction
- Date Received
- November 12, 2020
- Date of Event
- November 5, 2020
- Report Date
- January 28, 2021
- Manufacturer
- APPLIED MEDICAL RESOURCES
- Product Code
- NWV
- UDI-DI
- 00607915116866
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
THE EVENT UNIT WAS RETURNED TO APPLIED MEDICAL FOR EVALUATION. ENGINEERING DISASSEMBLED THE EVENT UNIT AND OBSERVED THAT ONE OF THE JAWS HAD DISLODGED FROM ITS TRACK. DAMAGE WAS ALSO OBSERVED ON THE POST OF THE JAW. BASED ON THE CONDITION OF THE RETURNED UNIT AND THE DESCRIPTION OF THE EVENT, THE JAWS WERE UNABLE TO OPEN DUE TO THE DISLODGEMENT OF ONE OF THE JAWS. IT IS LIKELY THAT THIS WAS CAUSED BY A LARGE FORCE THAT WAS APPLIED TO THE JAWS DURING USE, WHICH DAMAGED THE POST OF THE JAW. THE PROBABILITY AND CRITICALITY OF HARM RESULTING FROM THIS FAILURE HAVE BEEN EVALUATED AND WERE FOUND TO BE AT AN ACCEPTABLE LEVEL. THIS EVENT WAS INITIALLY REPORTED IN ERROR AND IS NOT REPORTABLE AS THE JAWS WERE REMOVED FROM THE TISSUE WITHOUT INJURY AND IT IS UNLIKELY TO CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY. THIS REPORT IS TO FOLLOW UP MEDWATCH REPORT # 2027111-2020-00616.
PROCEDURE PERFORMED: TLH GRASPER WOULD NOT RELEASE TISSUE AFTER HOLDING TISSUE BETWEEN JAWS THE CLINICIANS TRIED TO OPEN THE GRASPER BY REMOVING THE HANDLE AND THE JAWS STILL DIDN'T OPEN. THIS WAS ATTEMPTED 3 TIMES WITH AND WITHOUT THE HANDLE AND EVENTUALLY THE TISSUE WAS RELEASED WITH NO INJURY TO PATIENT. PATIENT STATUS: PATIENT IS FINE, AND COMPLETELY UNAFFECTED.
THE INCIDENT DEVICE IS ANTICIPATED TO RETURN. A FOLLOW-UP REPORT WILL BE PROVIDED UPON COMPLETION OF INVESTIGATION.
PROCEDURE PERFORMED: TLH. CER2020-002610 1 OF 2; CER2020-002611 2 OF 2. GRASPER WOULD NOT RELEASE TISSUE AFTER HOLDING TISSUE BETWEEN JAWS. THE CLINICIANS TRIED TO OPEN THE GRASPER BY REMOVING THE HANDLE, AND THE JAWS STILL DIDN'T OPEN. THIS WAS ATTEMPTED 3 TIMES WITH, AND WITHOUT THE HANDLE, AND EVENTUALLY THE TISSUE WAS RELEASED WITH NO INJURY TO PATIENT. INTERVENTION: NI. PATIENT STATUS: PATIENT IS FINE, AND COMPLETELY UNAFFECTED.