Narratives
EVALUATION SUMMARY: BASED ON THE ANALYSIS RESULTS, THIS COMPLAINT IS NOW DETERMINED TO BE A MDR MALFUNCTION. THE ANALYSIS SITE CONFIRMED THE CUSTOMER COMPLAINT AND FOUND THE UNIT DISPLAYED MULTIPLE GREEN AND BLUE CABLE ERRORS DUE TO THE TRANSLATIONAL AND ROTATIONAL CABLES. TO CORRECT THE CUSTOMER COMPLAINT, THE ANALYSIS SITE REPLACED THE TRANSLATIONAL AND ROTATIONAL CABLE. THE PORT SHAFT AND COIL PIN WAS REPLACED DUE TO 'IT WAS BENT', THE SAFETY LATCH WAS REPLACED DUE TO 'IT WAS WORN', AND THE SPUR GEAR WAS REPLACED DUE TO THE TEETH ON THE SPUR GEAR BEING WORN. HEAVY CORROSION ON THE UNIT CAUSED THE FIRING MECHANISM AND THE MANUAL SPADE ASSEMBLY TO RUST. TO CORRECT THIS ISSUE, THE ANALYSIS SITE REPLACED THE FIRING MECHANISM AND THE MANUAL SPADE ASSEMBLY. THE E-CLIP WAS REPLACED DUE TO 'IT WAS DAMAGED DURING DISASSEMBLY'. THE DEVICE HISTORY RECORD HAS BEEN REVIEWED AND HAS PASSED QA INSPECTION. COMPLAINT INFO IS TRENDED ON ARE REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.
IT WAS REPORTED THAT THE HOLSTER WAS GIVING A GREEN CABLE ERROR CODE DURING SAMPLING MODE DURING A BREAST BIOPSY. THE CABLING WASN'T TWISTED/BRAIDED ON ITSELF AND THE CONNECTIONS WERE SEATED TIGHTLY BUT IT WAS NOTICED THE GREEN CABLING WAS FRAYED. THE ERROR WAS BYPASSED AND THE CASE WAS COMPLETED WITH NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MTOME ST HOLSTER/CABLES | KNW | ETHICON ENDO-SURGERY, LLC | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CONTROL MODULE| PROBE |