FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1083021 · Received July 23, 2008

Report

Report Number
3005075853-2008-00645
Event Type
Malfunction
Date Received
July 23, 2008
Report Date
June 18, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE ANALYSIS WAS PERFORMED AND WAS FOUND THAT THE HANDPIECE NO LONGER MEETS THE SPECIFICATIONS FOR CONTINUITY. THE INSTRUMENT WAS TESTED ON THE GENERATOR AND GAVE AN ERROR CODE 3. THE HANDPIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND MOISTURE INGRESS WERE FOUND. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED WITH NO INFO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE LFL ETHICON ENDO-SURGERY, LLC NA NA

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR