FDA Adverse Event
Malfunction
Summary report: N
ULTRACISION HARMONIC SCALPEL HAND PIECE
MDR report key: 1083021
·
Received July 23, 2008
Report
- Report Number
- 3005075853-2008-00645
- Event Type
- Malfunction
- Date Received
- July 23, 2008
- Report Date
- June 18, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE ANALYSIS WAS PERFORMED AND WAS FOUND THAT THE HANDPIECE NO LONGER MEETS THE SPECIFICATIONS FOR CONTINUITY. THE INSTRUMENT WAS TESTED ON THE GENERATOR AND GAVE AN ERROR CODE 3. THE HANDPIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND MOISTURE INGRESS WERE FOUND. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED WITH NO INFO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC SCALPEL HAND PIECE | LFL | ETHICON ENDO-SURGERY, LLC | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |