FDA Adverse Event
Malfunction
Summary report: N
DIMENSION CLINICAL CHEMISTRY SYSTEM
MDR report key: 1082902
·
Received July 25, 2008
Report
- Report Number
- 1226181-2008-00085
- Event Type
- Malfunction
- Date Received
- July 25, 2008
- Date of Event
- June 30, 2008
- Report Date
- July 1, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC
- Product Code
- MMI
- PMA / PMN Number
- K944093
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A SIEMENS HEALTHCARE DIAGNOSTICS INC FIELD SERVICE REPRESENTATIVE (FSE) WAS DISPATCHED TO THE ACCOUNT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS CARRY IN FROM THE SAMPLE DRAIN. THE FSE CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
Description of Event or Problem · 1
A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | MMI | SIEMENS HEALTHCARE DIAGNOSTICS, INC | RXLMAX | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR |