DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT HAS BEEN COMPLETED. THE PROBLEM WAS CONFIRMED. UPON FURTHER INSPECTION, THE ELECTRODE BELT HAD PINS THAT WERE BENT INSIDE THE CONNECTOR. THE ROOT CAUSE OF THE BENT PINS WAS PROBABLY EXCESSIVE FORCE APPLIED TO THE CONNECTOR DURING MATING. THE CONNECTOR WAS FORCED INTO THE MONITOR WHICH DEFEATED THE CONNECTOR KEYING MECHANISM AND ALLOWED THE PINS TO MISALIGN WITH THE MATING SOCKETS. THE DAMAGED ELECTRODE BELT CONNECTOR WAS REPLACED. THE ELECTRODE BELT WAS RETESTED AND THEN RESTOCKED. NO ADVERSE EVENT RUSTLED FROM THE BENT PINS. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
A MALE PATIENT CONTACTED LIFECOR CUSTOMER SUPPORT TO REPORT THAT THE E-BELT CONNECTOR WAS NOT CONNECTING PROPERLY. HE STATED THAT SOME OF THE PINS LOOKED LIKE THEY WERE MISSING. HE STATED THAT HE WAS ALSO RECEIVING "CHECK BELT" MESSAGES. SUPPORT SENT A PATIENT SERVICES REPRESENTATIVE (PSR) TO THE PATIENT TO EXCHANGE THE ELECTRODE BELT.