LIFEVEST WCD 3100 SYSTEM
Report
- Report Number
- 3002158293-2008-00331
- Event Type
- Malfunction
- Date Received
- July 25, 2008
- Date of Event
- July 7, 2008
- Report Date
- July 24, 2008
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE MONITOR WAS FOUND TO HAVE A DEFECTIVE BATTERY CONNECTOR. THE CONNECTOR PINS WERE BENT. THE CONNECTOR WAS REPAIRED. THE MONITOR WAS RETESTED AND THEN RESTOCKED. THE ROOT CAUSE OF THE BENT PINS IS UNKNOWN, BUT IS LIKELY DUE TO A BATTERY PACK BEING FORCED INTO A MONITOR WITH THE CONNECTORS MISALIGNED. THE DAMAGED CONNECTOR ON THE MONITOR WAS REPLACED. NO ADVERSE EVENTS RESULTED FROM THE DAMAGED MONITOR. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY PROBLEMS WITH IT.
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING OF MONITOR, WHICH WAS RETURNED FOR ROUTINE MAINTENANCE, IT WAS DISCOVERED THAT MONITOR HAD A SMASHED BATTERY CONNECTOR, AND WOULD NOT POWER UP. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY PROBLEMS WITH IT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |