FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 1082848 · Received July 25, 2008

Report

Report Number
3004464228-2008-00147
Event Type
Malfunction
Date Received
July 25, 2008
Date of Event
June 26, 2008
Report Date
June 26, 2008
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WILL NOT BE RETURNED SO NO EVALUATION WAS POSSIBLE. THE CUSTOMER FELT RESISTANCE DURING THE FILL PROCESS, WHICH INDICATES A PROBABLE PROBLEM WITH A RETAINER THAT ALLOWED A COMPONENT TO MOVE RESULTING IN DAMAGE WHICH PREVENTED THE PLUNGER FROM ADVANCING. THE USER WOULD HAVE BEEN AWARE OF A PROBLEM DURING THE FILL PROCESS. THERE IS RESISTANCE FELT IN FILLING THE POD WITH INSULIN AND A DISTINCT "CRACKLING" NOISE RESULTING FROM STRIPPING THE THREADS OF THE COMPONENT WHEN OVER-PRESSURIZED. THE OMNIPOD USER GUIDE STATES: "WARNING" NEVER USE A POD IF, DURING FILL, YOU DETECT ANY CRACKLING NOISE OR RESISTANCE WHILE DEPRESSING THE PLUNGER OF THE FILL SYRINGE. USING A POD WITH THESE CONDITIONS COULD RESULT IN UNDER-DELIVERY OF INSULIN." THE USER IS ALSO INSTRUCTED IN THE USER GUIDE TO FREQUENTLY MONITOR THEIR BG LEVELS. BY FOLLOWING THESE RECOMMENDATIONS, THE USER WOULD BECOME AWARE OF HIGH BG LEVELS AND COULD USE ANOTHER DEVICE OR BACKUP THERAPY IF REQUIRED.

Description of Event or Problem · 1

A CUSTOMER CALLED IN TO REPORT THAT SHE WAS EXPERIENCING HIGH BG WITH NO OCCLUSION AND NO ALARMS. AFTER SPEAKING WITH CUSTOMER, IT WAS DETERMINED THAT SHE WAS USING A POD THAT WAS "HARD TO FILL." SHE STATED SHE DID NOT KNOW ABOUT THE LABELING AND SHE WAS ADVISED NOT TO USE HARD TO FILL PODS. SHE HAD POD ON HER TIP. THE CUSTOMER WAS ABLE TO ACTIVATE A NEW POD. NO FURTHER PROBLEMS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L11740

Patients

Seq Age Sex Outcome Treatment
1 Other