FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS (50)

MDR report key: 1082831 · Received July 24, 2008

Report

Report Number
1826988-2008-00809
Event Type
Malfunction
Date Received
July 24, 2008
Date of Event
July 1, 2008
Report Date
July 1, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER STATED THAT SHE OPENED A NEW CARTON OF TEST STRIPS AND FOUND THE CAP OPEN ON THE BOTTLE OF STRIPS. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION, AS EXPOSURE TO MOISTURE CAN CAUSE HIGH TEST RESULTS. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE, LLC NOT PROVIDED 7KC3C09

Patients

Seq Age Sex Outcome Treatment
1 UNK