INFINITY
Report
- Report Number
- 1220063-2008-00024
- Event Type
- Malfunction
- Date Received
- July 23, 2008
- Date of Event
- June 23, 2007
- Report Date
- July 23, 2008
- Manufacturer
- DRAEGER MEDICAL SYSTEMS, INC. (IT/M)
- Product Code
- MHX
- Removal / Correction Number
- 2517967-11/16/06-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
OUR INITIAL EVAL HAS CONFIRMED THAT THE TOP COVER OF THE DEVICE ACTUALLY SEPARATED FROM THE BASE AS REPORTED. THIS IS A KNOWN SUPPLIER ISSUE AND WE HAVE IMPLEMENTED A SUPPLIER CORRECTIVE ACTION AND INITIATED A FIELD CORRECTION. THE DEVICE IDENTIFIED IN THIS INCIDENT DOES NOT FALL IN THE RANGE OF THE POTENTIALLY DEFECTIVE DEVICES OF THE FIELD CORRECTION NUMBER: 2517967-11/16/06-001-C. WE ARE CONSIDERING THIS AN ISOLATED INCIDENT DUE TO THE AGE OF THE DEVICE. HOWEVER, WE WILL CONTINUE MONITORING THE COMPLAINTS AND IF ADD'L COMPLAINTS OF THIS NATURE ARE REPORTED FOR THE DEVICES PRIOR TO THE RECALL TIMELINE, WE WILL REVISIT THE SCOPE OF THE RECALL AND TAKE APPROPRIATE ACTION.
IT WAS REPORTED THAT THE COVER OF AN INFINITY DOCKING STATION (IDS), MFG IN 2003, DETACHED FROM ITS BASE. (B) (4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITY | PATIENT MONITOR DOCKING STATION | MHX | DRAEGER MEDICAL SYSTEMS, INC. (IT/M) | 5206110 | 5463559559 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |