FDA Adverse Event Malfunction Summary report: N

INFINITY

MDR report key: 1082791 · Received July 23, 2008

Report

Report Number
1220063-2008-00024
Event Type
Malfunction
Date Received
July 23, 2008
Date of Event
June 23, 2007
Report Date
July 23, 2008
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC. (IT/M)
Product Code
MHX
Removal / Correction Number
2517967-11/16/06-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OUR INITIAL EVAL HAS CONFIRMED THAT THE TOP COVER OF THE DEVICE ACTUALLY SEPARATED FROM THE BASE AS REPORTED. THIS IS A KNOWN SUPPLIER ISSUE AND WE HAVE IMPLEMENTED A SUPPLIER CORRECTIVE ACTION AND INITIATED A FIELD CORRECTION. THE DEVICE IDENTIFIED IN THIS INCIDENT DOES NOT FALL IN THE RANGE OF THE POTENTIALLY DEFECTIVE DEVICES OF THE FIELD CORRECTION NUMBER: 2517967-11/16/06-001-C. WE ARE CONSIDERING THIS AN ISOLATED INCIDENT DUE TO THE AGE OF THE DEVICE. HOWEVER, WE WILL CONTINUE MONITORING THE COMPLAINTS AND IF ADD'L COMPLAINTS OF THIS NATURE ARE REPORTED FOR THE DEVICES PRIOR TO THE RECALL TIMELINE, WE WILL REVISIT THE SCOPE OF THE RECALL AND TAKE APPROPRIATE ACTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE COVER OF AN INFINITY DOCKING STATION (IDS), MFG IN 2003, DETACHED FROM ITS BASE. (B) (4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITY PATIENT MONITOR DOCKING STATION MHX DRAEGER MEDICAL SYSTEMS, INC. (IT/M) 5206110 5463559559

Patients

Seq Age Sex Outcome Treatment
1