FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR PLUS DEFIBRILLATOR

MDR report key: 1082703 · Received July 23, 2008

Report

Report Number
3015876-2008-00821
Event Type
Malfunction
Date Received
July 23, 2008
Date of Event
May 6, 2008
Report Date
June 24, 2008
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K011144
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. THE INTERNAL BATTERY WAS FOUND TO BE DEPLETED AND THE DEVICE WOULD NOT POWER ON WITH POWER SUPPLY. THE ROOT CAUSE OF THE REPORTED PROBLEM WAS DETERMINED TO BE DUE TO TIN WHISKER GROWTH ON THE ON/OFF SWITCH FLEX CABLE. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS DISPLAYING THE CHARGE-PAK ICON IN THE READINESS INDICATOR. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK CR PLUS DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC CRPLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA