FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK CR PLUS DEFIBRILLATOR
MDR report key: 1082703
·
Received July 23, 2008
Report
- Report Number
- 3015876-2008-00821
- Event Type
- Malfunction
- Date Received
- July 23, 2008
- Date of Event
- May 6, 2008
- Report Date
- June 24, 2008
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K011144
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. THE INTERNAL BATTERY WAS FOUND TO BE DEPLETED AND THE DEVICE WOULD NOT POWER ON WITH POWER SUPPLY. THE ROOT CAUSE OF THE REPORTED PROBLEM WAS DETERMINED TO BE DUE TO TIN WHISKER GROWTH ON THE ON/OFF SWITCH FLEX CABLE. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS DISPLAYING THE CHARGE-PAK ICON IN THE READINESS INDICATOR. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK CR PLUS DEFIBRILLATOR | MKJ | PHYSIO-CONTROL, INC | CRPLUS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |