8100 ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2020-37026
- Event Type
- Malfunction
- Date Received
- November 11, 2020
- Report Date
- June 7, 2018
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NURSE
Narratives
THE REPORTED EVENT OF IUI CORROSION FILE WAS OPENED TO DOCUMENT AN EVENT THAT THE CUSTOMER REPORTED. NO DEVICES OR LOGS WERE RETURNED FOR INVESTIGATION. NO FURTHER INVESTIGATION OF THIS EVENT IS POSSIBLE AT THIS TIME. ALTHOUGH NO DEVICES WERE PROVIDED FOR INVESTIGATION, THE SITE VISIT SUMMARY CONTAINS PHOTOS THAT CONFIRM THE CUSTOMER¿S GENERALIZED REPORT. AN INVESTIGATION CAN¿T BE PERFORMED USING THE ATTACHED PHOTO ALONE. NO DEVICE HISTORY OR QN SEARCH WAS PERFORMED SINCE NO SERIAL NUMBER FOR THE SUSPECT DEVICE WAS REPORTED BY THE CUSTOMER. A DHR (DEVICE HISTORY RECORD) REVIEW CANNOT BE COMPLETED AS THE SERIAL NUMBER WAS NOT OBTAINED UPON RECEIPT OF THE COMPLAINT. ADDITIONALLY, A HISTORICAL REVIEW OF COMPLAINTS IN TRACKWISE CANNOT BE CONDUCTED. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
THE CUSTOMER REPORTED DURING A CPC SITE VISIT CORRODED IUI'S. IT WAS FOUND DURING THE CPC SITE VISIT THAT THERE ARE TWO DEPARTMENTS RESPONSIBLE FOR CLEANING THE DEVICES. THE TRANSPORT TEAM AND THE EVS DEPARTMENT. THE EVS DEPARTMENT IS RESPONSIBLE FOR CLEANING THE DEVICES THAT ARE LOCATED IN A ROOM ONCE THE PATIENT HAS BEEN DISCHARGED, THE TRANSPORT TEAM CLEANS ALL OTHERS. IN DEMONSTRATION, THE TRANSPORT TEAM MANAGER CLEANED THE DEVICES WITH PDI BLEACH WIPES WHICH INCLUDED WIPING DIRECTLY OVER THE IUI CONNECTORS, THOROUGHLY, WIPING THE ENTIRE CASE, AND INSIDE THE LVP DOOR. THE PUMPS ARE THEN STORED IN THEIR DEPO WITH SOME OF THEM PLUGGED IN AND OTHERS NOT DUE TO LIMITED AC OUTLETS. THE EVS MANAGER STATED THAT THEY ALSO USE THE PDI BLEACH WIPES AND WIPE OVER THE ENTIRE CASE, INCLUDING THE IUI CONNECTORS. ONCE THE DEVICE HAS BEEN CLEANED, IT IS IMMEDIATELY BAGGED AND PLUGGED IN. IT WAS NOTED THAT NEITHER TEAM REMOVED THE CLEANING SOLUTION AFTER CLEANING AND THE BLEACH FILM WAS OBSERVED ON ALL DEVICES. THE BIOMED DEPARTMENT WAS VERY KNOWLEDGEABLE AND WAS AWARE OF THE BLEACH WIPES TO THE IUI CONNECTORS AND CONNECTED THIS TO THE FREQUENT CHANNEL DISCONNECTS. ON A SIDE NOTE, THE PCU'S ARE RENTED AND THE LVPS ARE OWNED BY THE FACILITY. ADDITIONALLY, THE DEVICES HAVE ONE LVP BOLTED TO THE RIGHT OF THE PCU AND NEITHER EVS OR THE TRANSPORT TEAM DETACH THE LVP FOR CLEANING PURPOSES. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1286565 | 8100 ALARIS PUMP MODULE | PUMP, INFUSION | FRN | CAREFUSION SD | 8100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |