FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE ARTHRITIS PAD
MDR report key: 1082657
·
Received July 21, 2008
Report
- Report Number
- 1811605-2008-00105
- Event Type
- Malfunction
- Date Received
- July 21, 2008
- Date of Event
- July 1, 2008
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THE FABRIC FOUNDATION OF THE UNIT WAS BURNED BY AN INTERNAL COMPONENT. INSPECTION OF THE UNIT REVEALED A 1/2" HOLE IN ONE SURFACE, ALTHOUGH THERE WAS NO EVIDENCE OF AN ACTUAL BURN. THE CONDITION OF THE UNIT SUGGESTED THAT THE UNIT HAD BEEN FOLDED DURING USE, WHICH HAD INSULATED AND INTENSIFIED THE HEAT LEVEL. THERE WAS ALSO EVIDENCE OF COMPRESSING AND DAMAGE TO INTERNAL COMPONENTS. WE CONCLUDED THAT THERE WAS NO ACTUAL DEFECT FOUND AND THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE ON THE PART OF THE CONSUMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE ARTHRITIS PAD | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT COMPANY | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |