FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE ARTHRITIS PAD
MDR report key: 1082655
·
Received July 21, 2008
Report
- Report Number
- 1811605-2008-00107
- Event Type
- Malfunction
- Date Received
- July 21, 2008
- Date of Event
- July 1, 2008
- Report Date
- July 14, 2008
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THE FABRIC FOUNDATION OF THE UNIT WAS BURNED BY AN INTERNAL COMPONENT. INSPECTION OF THE UNIT REVEALED A TERMINAL HAD BROKEN AT THE LEAD WIRE. THE CONDITION OF THE UNIT SUGGESTED THAT THE UNIT HAD BEEN SEVERELY COMPRESSED, WITH DAMAGE TO MANY INTERNAL COMPONENTS. WE CONCLUDED THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE ON THE PART OF THE CONSUMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE ARTHRITIS PAD | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT COMPANY | 055 |