FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER EDTA 2K

MDR report key: 10826416 · Received November 11, 2020

Report

Report Number
1917413-2020-01138
Event Type
Malfunction
Date Received
November 11, 2020
Date of Event
October 14, 2020
Report Date
December 2, 2020
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
PMA / PMN Number
K981013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD RECEIVED SAMPLES AND 3 PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR EMBEDDED FM WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, THE CUSTOMER SAMPLES WERE EVALUATED BY VISUAL EXAMINATION AND THE INDICATED FAILURE MODE FOR EMBEDDED FM WITH THE INCIDENT LOT WAS OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED THE BD VACUTAINER® EDTA 2K HAD FOREIGN MATTER IN TUBE; BIOLOGICAL AND NON-BIOLOGICAL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER STATED "THE FOLLOWING ISSUES WERE FOUND IN TUBES (367846) FROM LOT 0100257: SPOT IN TUBE X2 DIRTY TUBE X1

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE BD VACUTAINER® EDTA 2K HAD FOREIGN MATTER IN TUBE; BIOLOGICAL AND NON-BIOLOGICAL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER STATED "THE FOLLOWING ISSUES WERE FOUND IN TUBES (367846) FROM LOT 0100257: SPOT IN TUBE X2, DIRTY TUBE X1.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1287355 BD VACUTAINER EDTA 2K BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 0100257

Patients

Seq Age Sex Outcome Treatment
1