FDA Adverse Event Malfunction Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL

MDR report key: 1082602 · Received July 23, 2008

Report

Report Number
2023826-2008-00997
Event Type
Malfunction
Date Received
July 23, 2008
Date of Event
June 27, 2008
Report Date
July 11, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P880091
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON WAS INSERTING A ONE PIECE TORIC SILICONE LENS MODEL AA4203TL AND THE LENS TORE. THE SURGEON CUT THE LENS INTO PIECES TO REMOVE IT, AND THEY DISCARDED THE LENS IN OR. THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL CO. AA4203TL N/A

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE MODEL: MTC-60CFP| INJECTOR MODEL: MSI-PR