FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC MOIST HEAT PACK

MDR report key: 1082561 · Received July 21, 2008

Report

Report Number
1811605-2008-00111
Event Type
Malfunction
Date Received
July 21, 2008
Date of Event
June 1, 2008
Report Date
July 7, 2008
Manufacturer
BATTLE CREEK EQUIPMENT COMPANY
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THE UNIT BURNED. VISUAL INSPECTION OF THE UNIT REVEALED THAT SEVERAL INTERNAL COMPONENTS WERE BENT AND A SEGMENT OF THE LEAD WIRE HAD BROKEN IN ONE AREA, ALLOWING A SPARK TO CONTACT THE FABRIC FOUNDATION AND CAUSING A BURN. WE DETERMINED THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE OF THE UNIT ON THE PART OF THE CONSUMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE AUTOMATIC MOIST HEAT PACK ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT COMPANY 055

Patients

Seq Age Sex Outcome Treatment
1