FDA Adverse Event
Malfunction
Summary report: N
STERRAD NX STERILIZER
MDR report key: 1082512
·
Received July 24, 2008
Report
- Report Number
- 2084725-2008-00399
- Event Type
- Malfunction
- Date Received
- July 24, 2008
- Date of Event
- July 2, 2008
- Report Date
- July 2, 2008
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K042116
- Removal / Correction Number
- Z-1631-2008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(OIL MIST) - CAPITAL EQUIPMENT, UNIT EVALUATED AT THE FACILITY. THE FSE FOUND OIL PLUG ON VACUUM PUMP HAD HOLE CAUSING MISTING ISSUE. THE FSE SENT THE CUSTOMER REPLACEMENT OIL PLUGS TO CORRECT PROBLEM. THIS ISSUE IS BEING ADDRESSED WITH A CUSTOMER LETTER, RELATED TO CORRECTION & REMOVAL.
Description of Event or Problem · 1
THE CUSTOMER REPORTED "OIL MIST HAZE". THE CUSTOMER STATED THAT THEY WERE NO PHYSICAL COMPLAINTS REPORTED. THE ASP FIELD SERVICE ENGINEER (FSE) WENT TO THE FACILITY TO REPAIR THE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STERRAD NX STERILIZER | STERRAD EQUIPMENT (MLR) | MLR | ADVANCED STERILIZATION PRODUCTS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |