FDA Adverse Event Malfunction Summary report: N

STERRAD NX STERILIZER

MDR report key: 1082512 · Received July 24, 2008

Report

Report Number
2084725-2008-00399
Event Type
Malfunction
Date Received
July 24, 2008
Date of Event
July 2, 2008
Report Date
July 2, 2008
Manufacturer
ADVANCED STERILIZATION PRODUCTS
Product Code
MLR
PMA / PMN Number
K042116
Removal / Correction Number
Z-1631-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(OIL MIST) - CAPITAL EQUIPMENT, UNIT EVALUATED AT THE FACILITY. THE FSE FOUND OIL PLUG ON VACUUM PUMP HAD HOLE CAUSING MISTING ISSUE. THE FSE SENT THE CUSTOMER REPLACEMENT OIL PLUGS TO CORRECT PROBLEM. THIS ISSUE IS BEING ADDRESSED WITH A CUSTOMER LETTER, RELATED TO CORRECTION & REMOVAL.

Description of Event or Problem · 1

THE CUSTOMER REPORTED "OIL MIST HAZE". THE CUSTOMER STATED THAT THEY WERE NO PHYSICAL COMPLAINTS REPORTED. THE ASP FIELD SERVICE ENGINEER (FSE) WENT TO THE FACILITY TO REPAIR THE UNIT.