FDA Adverse Event Malfunction Summary report: N

CAPTIVE TWIST STRAIGHT DRIVER

MDR report key: 1082424 · Received July 21, 2008

Report

Report Number
2249697-2008-00208
Event Type
Malfunction
Date Received
July 21, 2008
Date of Event
June 30, 2008
Report Date
June 30, 2008
Manufacturer
STRYKER ORTHOPAEDICS MAHWAH
Product Code
HXX
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RESULT OF THE INVESTIGATION INDICATES THAT THE CAPTIVE TWIST STRAIGHT DRIVER WAS USED TO TIGHTEN THE SCREW. AS A RESULT, THE DRIVER TIP WAS TORSIONALLY LOADED CAUSING DRIVER TIP TO FRACTURE IN TORSIONAL SHEAR. IT SHOULD BE NOTED THAT THE CAPTIVE TWIST DRIVER IS ONLY INTENDED TO INITIATE A SCREW OR PLUG INTO THE SHELL. ONLY A STANDARD T-20 DRIVER, CAT 2107-101X, SHOULD BE USED TO TORQUE A SCREW OR PLUG INTO THE SHELL. IF ADDITIONAL INFO BECOMES AVAILABLE A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT "THE CAPTIVE TWIST TIP BROKE OFF IN THE SCREWHEAD OF THE CAPSCREW IN THE TRIDENT SHELL. TIP COULD NOT BE REMOVED, LEFT IN PT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPTIVE TWIST STRAIGHT DRIVER INSTRUMENT HXX STRYKER ORTHOPAEDICS MAHWAH NA S2C06

Patients

Seq Age Sex Outcome Treatment
1 NI Other