FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT

MDR report key: 1082386 · Received July 25, 2008

Report

Report Number
2183996-2008-01101
Event Type
Malfunction
Date Received
July 25, 2008
Date of Event
July 15, 2008
Report Date
July 15, 2008
Manufacturer
DISETRONIC MEDICAL SYSTEMS
Product Code
LZG
PMA / PMN Number
K060876
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

IN 2008, THE PATIENT REPORTED RECEIVING AN E4 (OCCLUSION) ERROR ON HIS INFUSION DEVICE. TO TROUBLESHOOT THE PATIENT WAS INSTRUCTED TO DISCONNECT FROM HIS INFUSION SITE AND TO PERFORM A PRIME. HE STATED THAT HE HAD ATTEMPTED TO DO THIS BEFORE THE REPORT BUT HE RECEIVED AN E10 (CARTRIDGE) ERROR. THE PATIENT WAS INSTRUCTED TO REMOVE THE INSULIN CARTRIDGE AND TO PERFORM THE CHANGE CARTRIDGE PROCEDURE. THE PATIENT STATED THAT HE PRESSED AND HELD THE CHECK BUTTON FOR 3 SECONDS AND THE PISTON ROD EXTENDED INSTEAD OF RETRACTED. HE THEN RECEIVED AN E10 ERROR. THE PATIENT WAS INSTRUCTED TO SWITCH TO HIS BACKUP INFUSION DEVICE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. PRODUCT WAS REPLACED AND REQUESTED TO BE RETURNED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCU-CHEK SPIRIT INSULIN INFUSION PUMP LZG DISETRONIC MEDICAL SYSTEMS NA NA

Patients

Seq Age Sex Outcome Treatment
1 INSULIN INFUSION SET| INSULIN